ARRAY_TESTS uses eight frames and derives both the STARE and ND_STARE readout noises, which are both filed with the calibration system. The range of acceptable dark current is 0-10 electrons/sec compared with -1-+1 for UFTI, and the nominal ND_STARE readout noise is 20-50 electrons compared with 8-30 electrons for UFTI. The STARE mode readout noise should be 30-70 electrons to be nominal. The log file goes to $ORAC_DATA_OUT/ircam3_array_tests.log.